LI Qiang, WANG Zhongyu, WANG Zhuoranet al. Uncertainty evaluation for the dynamic calibration of pressure transducer[J]. Journal of Beijing University of Aeronautics and Astronautics, 2015, 41(5): 847-856. doi: 10.13700/j.bh.1001-5965.2014.0356(in Chinese)
Citation: LI Qiang, WANG Zhongyu, WANG Zhuoranet al. Uncertainty evaluation for the dynamic calibration of pressure transducer[J]. Journal of Beijing University of Aeronautics and Astronautics, 2015, 41(5): 847-856. doi: 10.13700/j.bh.1001-5965.2014.0356(in Chinese)

Uncertainty evaluation for the dynamic calibration of pressure transducer

doi: 10.13700/j.bh.1001-5965.2014.0356
  • Received Date: 16 Jun 2014
  • Rev Recd Date: 25 Aug 2014
  • Publish Date: 20 May 2015
  • The uncertainty of dynamic calibration for pressure transducer is a key indicator of its dynamic measurement precision. A novel grey method (GM) was proposed to evaluate the dynamic calibration uncertainty of pressure transducer. Firstly, a sine pressure generator was used to produce required sine pressure signal to drive the calibrated pressure transducer and obtain output of transducer. Secondly, using the grey relational analysis to deal with the measurement sequences outputted, a weighted value sequence was obtained. Thirdly, a grey model was established to calculate the calibration uncertainty of appointed frequency. Finally, a weighted least-square fitting method was applied to do fitting with the calibration uncertainty of appointed frequency, and then a dynamic calibration uncertainty model was established. An experimental verification under the calibration condition of changing amplitude and unchanging amplitude was set up. Grey method and Huang method (HM) were used to process the output of pressure transducer separately. The results show that the curve models of dynamic calibration uncertainty are consistent with each other in the setting condition. What is more, under changing amplitude, the relative error of calibration uncertainty obtained by the two methods is superior to 6% at appointed frequency points and is also less-than 10% at testing frequency points. Under unchanging amplitude, the relative error is less than 5% at most of frequency points and is superior to 0.018% at some testing frequency points. Based on the above analysis, it can be approved that the grey method can evaluate the dynamic calibration uncertainty of pressure transducer reliably.

     

  • loading
  • [1]
    Zheng D Z,Wu Z L,Yan Q,et al.Dynamic performance improvement on resonant vibration cylinder pressure transducer[J].Advanced Science Letters.2011,4(4-5):1376-1380.
    [2]
    Sharifian S A,Buttsworth D R.Evaluation of glued-diaphragm fibre optic pressure sensors in a shock tube[J].Shock Waves,2007,16(3):189-197.
    [3]
    Skoglund M, Pettersson M.A calibration scheme for imperfect quantizers[J].IEEE Transactions on Instrumentation and Measurement,2000,49(5):1063-1068.
    [4]
    Medawar S,Handel P,Murmann B,et al.Dynamic calibration of undersampled pipelined ADCs by frequency domain filtering[J].IEEE Transactions on Instrumentation and Measurement,2013,62(7):1882-1891.
    [5]
    黄俊钦. 动态不确定度的估算方法和应用实例[J].计量学报,2005,26(4):372-375. Huang J Q.The estimation method and application of dynamic uncertainty[J].Acta Metrologica Sinica,2005,26(4):372-375(in Chinese).
    [6]
    黄俊钦. 测温系统动态校准结果的不确定度估[J].仪器仪表学报,2008,29(4):870-873. Huang J Q.Dynamic uncertainty estimation of dynamic calibration results in temperature measurement system[J].Chinese Journal of Scientific Instrument,2008,29(4):870-873(in Chinese).
    [7]
    Zhang Y,Zu J,Zhang H Y.Dynamic calibration method of high-pressure transducer based on quasi-delta function excitation source[J].Measurement,2012,45(8):1981-1988.
    [8]
    Martz Jr H F,Waller R A. Bayessian zero-failure (BAZE) reliability demonstration test procedure[J]. Journal of Quality Technology,1979,11(3):128-137.
    [9]
    Crowder S V,Moyer R D.A two-stage Monte Carlo approach to the expression of uncertainty with non-linear measurement equation and small sample size[J].Metrologia,2006,43(1):34-41.
    [10]
    Miura T,Huete A R,Yoshioka H.Evaluation of sensor calibration uncertainties on vegetation indices for MODIS[J].IEEE Transactions on Geoscience and Remote Sensing,2000,38(3):1399-1409.
    [11]
    Yadav S,Kumaraswamy B V,Gupta V K,et al.Least squares best fit line method for the evaluation of measurement uncertainty with electromechanical transducers (EMT) with electrical outputs (EO)[J].Mapan Journal of Metrology Society of India,2010,25(2):97-106.
    [12]
    Xia X T,Wang Z Y,Gao Y S.Estimation of non-statistical uncertainty using fuzzy-set theory[J].Measurement Science and Technology,2000,11(4):430-435.
    [13]
    Xia X T,Chen X Y,Zhang Y Z,et al.Grey bootstrap method of evaluation of uncertainty in dynamic measurement[J].Measurement,2008,41(6):687-696.
    [14]
    王中宇,刘智敏,夏新涛.测量误差与不确定度评定[M].北京:科学出版社,2008:35-40. Wang Z Y,Liu Z M,Xia X T.Measurement error and uncertainty evaluation[M].Beijing:Science Press,2008:35-40(in Chinese).
    [15]
    Wang Z Y,Ge L Y.Novel method of evaluating dynamic repeated measurement uncertainty[J].Journal of Testing and Evaluation,2008,36(5):453-459.
    [16]
    Qin P,Shen Y,Wang Z Y.Grey evaluation of non-statistical uncertainty in multidimensional precision measurement[J].The International Journal of Advanced Manufacturing Technology,2006,31(5-6):539-545.
    [17]
    费业泰. 误差理论与数据处理[M].北京:机械工业出版社,2005:23-26. Fei Y T.Error theory and data processing[M].Beijing:Machinery Industry Press,2005:23-26(in Chinese).
    [18]
    盛克仁. 过程测量仪表[M].北京:化学工业出版社,1992:15-18. Sheng K R.Process measuring instruments[M].Beijing:Chemical Industry Press,1992:15-18(in Chinese).
    [19]
    袁俊先,蔡菁. 基于正弦压力校准装置的压力传感器不确定度评定[J].计测技术,2013(S1):190-192. Yuan J X,Cai J.Uncetainty evaluation for pressure transducer based on the sine pressure generator[J].Testing Technology,2013(S1):190-192(in Chinese).
    [20]
    张力. 正弦压力校准测量不确定度分析[C]//压力计量服务和测试技术研讨会论文集.北京:中国计量测试学会,2003:26-32. Zhang L.Evaluation of uncertainty of sinusoidal pressure calibration[C]//Pressure Measurement Services and Test Technology Seminar Proceedings.Beijing:Chinese Society for Measurement,2003:26-32(in Chinese).
    [21]
    JJG 624—2005 动态压力传感器检定规程[S].北京:机械工业出版社,2005:5-9. JJG 624—2005 Verification regulation of dynamic pressure transducers[S].Beijing:Machinery Industry Press,2005:5-9(in Chinese).
    [22]
    刘思峰,党耀国,方志耕,等.灰色系统:理论与应用[M].北京:科学出版社,2010:5-10. Liu S F,Dang Y G,Fang Z G,et al.The grey system: theory and its application[M].Beijing:Science Press,2010:5-10(in Chinese).
    [23]
    Liu S F. On measure of grey information[J].The Journal of Grey System,1995,7(2):97-101.
    [24]
    Shu W. A study of the dynamic measurement uncertainty[J].Acta Metrologia Sinica,2003,24:117-122.

  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views(1143) PDF downloads(543) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return