Volume 31 Issue 07
Jul.  2005
Turn off MathJax
Article Contents
Hu Yida, Li Dazhai, Yang Yang, et al. Integration of microsensor for microsurgery robot's end-effector[J]. Journal of Beijing University of Aeronautics and Astronautics, 2005, 31(07): 794-798. (in Chinese)
Citation: Hu Yida, Li Dazhai, Yang Yang, et al. Integration of microsensor for microsurgery robot's end-effector[J]. Journal of Beijing University of Aeronautics and Astronautics, 2005, 31(07): 794-798. (in Chinese)

Integration of microsensor for microsurgery robot's end-effector

  • Received Date: 30 Oct 2004
  • Publish Date: 31 Jul 2005
  • In order to enhance the effect of robotic microsurgery, the microsensors were integrated on the robot’s end-effector. On the basis of requirements presented for the integration design, measuring mechanism for the robotic end trephine’s force and cutting depth was studied. Force microsensor and position microsensor were made use of to measure the surgical information of force and depth. The measuring mechanism was practised by means of linear sliding bearing and differential measuring structure. The sensor data board was developed. With the power spectral estimation of sensor data, the two digital filtering methods were proposed to eliminate the interference to the original microsensor signal. They are a filtering method of lowpass-bandstop serial structure suitable for the PC and a shift average filtering arithmetic suitable for the sensor data board. The experimental results show the integration of microsensors for microsurgery robot’s end-effector is qualified for the design requirements, and the robotic end trephine can fulfill the surgical task of corneal cutting accurately.

     

  • loading
  • [1] Howe R D, Matsuoka Y. Robotics for surgery[J] Annu Rev Biomed Eng, 1999,1:211~240 [2] Salcudean S E, Yan J. Towards a force-reflecting motion-scale system for microsurgery. In:Proceedings of the IEEE International Conference on Robotics and Automation. San Diego, 1994.2296~2301 [3] Kwon D S, Woo K Y, Song S K, et al. Microsurgical telerobot system. In:Proceedings of IEEE/RSJ International Conference on Intelligent Robots and Systems. Victoria, 1998.13~17 [4] 奥本海姆A V,谢弗R W.离散时间信号处理[M] 黄建国, 刘树棠译.北京:科学出版社,1998.400 Oppenheim A V, Schafer R W. Discrete-time signal processing[M] Translated by Huang Jianguo, Liu Shutang. Beijing:Science Press, 1998.400
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views(2657) PDF downloads(987) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return