Experiments on Performance of a Molecular Sieve Oxygen Concentrator System

  • Received Date: 01 Aug 2001
  • Publish Date: 31 Jan 2003
  • The prototype of a molecular sieve oxygen concentrator system (MSOCS) was developed, on the principle of PAS. Various intake pressures,output flow of MSOCS, altitude pressure and the environment temperature tests were tested to analyze its performance of generating oxygen. The results showed that the concentration of oxygen generation of the MSOCS increases with increasing the intake pressures and altitude, and decreases with increasing output flow of MSOCS. The gradient rate and the environment temperature make no effects. The performance of generating oxygen meets the physiological requirements for the system engineering.

     

  • loading
  • [1] 肖华军,袁修干.机载分子筛制氧技术发展的现状与动向[J] 空科学技术,1997,(1):26~28 Xiao Huajun,Yuna Xiugan. Current state and trend of molecular sieve oxygen generating system in technology[J] Aeronautics Science and Technology, 1997,(1):26~28(in Chinese) [2] 肖华军.英国机载分子筛制氧系统发展现状[J] 空与航天,1994,(1):47~51 Xiao Huajun. The current development of Britain on molecular sieve oxygen generating system[J] Aeronautics & Astronautics, 1994,(1):47~51(in Chinese) [3] 肖华军,袁修干.飞行环境综合因素对飞机氧气装备供氧参数的影响[J] 京航空航天大学学报,1997,23(5):586~589 Xiao Huajun,Yuan Xiugan. Effects of inflight environment multiple factors on the parameter of aircraft oxygen equipment[J] Journal of Beijing University of Aeronautics and Astronautics, 1995,23(5):586~589(in Chinese) [4] 肖华军.法国军用飞机机载分子筛氧气系统[J] 航空与航天,1996,(1):38~40 Xiao Huajun. The molecular sieve oxygen generating system of France' military aircraft[J] Aeronautics & Astronautics, 1996,(1):38~40(in Chinese) [5] Ikels K G.Performance criteria for the MOSOGS .Proceedings of the 25th Annual Symposium .Las Vegas,Nevada:SAFE Association,1987.29 [6] Xiao Huajun. The physiological requirement on the concentration of aircrafts' oxygen supply equipment .Proceedings of 29th SAFE Annual Symposium .Las Vegas,Nevada:SAFE Association,1991.192